- Series
- UV / Visible
- Accuracy (wavelenght)
- ± 0,3
- Bandwidth
- 1 nm
- Connection
- USB-A / USB-B / RS-232 / Ethernet
- Dimension
- 456 x 360 x 185
- Basic photometry
- Yes
- Frequency
- 50/60
- File Management
- Yes
- Communication interfaces
- Touch screen
- Selectable languages
- IT - EN - CHN - ESP - DEU - FRA - POR - RUS
- Diffused light
- ≤ 0,03 %T @ 220 e 360 nm
- Memory
- 64 GB (internal) expandable with USB stick and SD card
- Weight
- 17,0
- Quantitative
- Yes
- Photometric range % T
- 0 - 400 %T
- Photometric range Abs
- -4,0 - +4,0 Abs
- Photometric range Conc
- 0 - 9999,9 Conc
- Wavelenght range
- 190 - 1100 nm
- Range (UV/Visible)
- 190 - 1100 nm
- Photometric resolution
- 0,001
- Wavelength resolution
- 0,1 nm
- Spectral scanning
- Yes
- Standard cell support
- Fixed 1 position, 10 mm path lenght
- System Utilities
- Yes
- Sample compartment
- Accessible from above and the front
- Power supply
- 100 - 240
- Performance verification
- Yes
- Power
- 140
- Optical system
- Double beam
- Sources
- Tungsten-halogen/Deuterium
- Stability
- ±0,002 Abs/hr (500 nm after 2 hours warm-up)
- Accuracy (photometric)
- ± 0,002 Abs (0,0 - 0,5 Abs)
± 0,004 Abs (0,5 - 1,0 Abs)
± 0,3 %T (0,0 - 100,0 %T)
- Display
- Colour LCD touch screen display 10.1'', 1280 x 1280
- Detector
- Silicon photodiodes
- WL Repeteability
- ≤ 0,2 nm
- Rotation speed
- 6000 nm/min
- Scan speed
- 20 - 3200 nm/min
- Photometric repeteability
- ≤ 0,001 Abs (0,0 - 0,5 Abs)
≤ 0,002 Abs (0,5 - 1,0 Abs)
≤ 0,15 %T (0 - 100 %T)
- Noise
- ≤ 0,0002 Abs (at 0,0 Abs) @ 500 nm
≤ 0,0004 Abs (at 1,0 Abs) @ 500 nm
≤ 0,0008 Abs (at 2,0 Abs) @ 500 nm
- Baseline drift
- ± 0,0005 Abs
- Functions
- Photometry, Quantitative, Multi WL, Spectrum scan, Time scanning, Kinetics, DNA/Protein, Custom methods, File management, System, Performance verification