Spectrophotometers

ONDA V-10 PLUS Spectrophotometer (VIS)

ONDA V-10 PLUS Spectrophotometer (VIS)

Features

VISIBLE Single beam

Range 325 - 1000 nm

Bandwith 4 nm

Other Features

VISIBLE single-beam spectrophotometer

Single beam optical system, 4 nm bandwidth, Tungsten source

LCD display, membrane keypad

Spectral resolution 0.1 nm

Photometric accuracy ±0.5%T

Photometric accuracy ±0.3%T

Photometric range 0...200%T, -0.3...3 Abs, 0...9999,9 Conc

Functions: Photometry, Concentration, Coefficients

External dimensions (WxDxH) 490x360x210 mm

Weight 12 kg

Power supply 100 - 240 V AC, 50/60 Hz

Power 75 W

Included Accessories

-4-position holder for 10 mm cuvettes -4 optical glass cells, 10 mm optical path -Power cable, dust cover -Calibration report -MWave Basic software (for quantitative, single λ photometric, kinetic analysis)
Product Code
11000052
Category
Spectrophotometers
Series
Visible
Accuracy (wavelenght)
± 2 nm
Bandwidth
4 nm
Connection
USB-B / RS-232
Dimension
490 x 360 x 210
Basic photometry
Yes
Warranty
2 years
File Management
Yes
Communication interfaces
Membrane keyboard
Selectable languages
Inglese
Diffused light
< 0,3 %T
Memory
200 data sets
Weight
12 kg
Quantitative
Yes
Photometric range % T
0 ... 200 %T
Photometric range Abs
-0,3 ... 3 Abs
Photometric range Conc
0 ... 9999,9 Conc
Wavelenght range
325 - 1000 nm
Range (UV/Visible)
Visibile
Photometric resolution
0,001
Wavelength resolution
0,1 nm
Standard cell support
Manual 4 positions, cuv. optical path 10 mm
Sample compartment
Accessible from above
Power supply
100 - 240 V, 50 / 60 Hz
Power
75 W
Optical system
Single beam
Source of light
Tungsten - halogen
Stability
± 0.002 Abs/hr (at 500 nm after 2 hours of warm-up)
Accuracy (photometric)
±0,5 (0-100) %T
Display
LCD display 128 x 64 px
Detector
Silicon photodiodes
WL Repeteability
± 0,8 nm
Photometric repeteability
±0,3 (0-100) %T
Functions
Photometry, Quantitative, Coefficients